A Simplified Approach for Real-Time Detection of Arterial Wall Velocity and Distension

Piero Tortoli, Rossano Bettarini, Francesco Guidi, Fabio Andreuccetti, and Daniele Righi

ABSTRACT Arterial stiffness is known to increase with age and with many vascular diseases, but its noninvasive assessment in patients still represents a difficult task. The measurement of diameter change during the cardiac cycle (distension) has been proposed as a means to estimate arterial compliance and stiffness. Therefore, we have developed a simple PC-based device and algorithm for noninvasive quantification of vessel wall motion and diameter change in humans. This goal is achieved in real-time by processing the base-band signals from a commercial ultrasound Doppler system. Real-time operation is of crucial importance, because it allows a rapid achievement of optimal measurement conditions.

The system was evaluated in a laboratory using a string phantom and was tested on the carotid arteries of 10 volunteers. Wall velocities from 0.05 to 600 mm/s and displacements lower than 2 μm were detected with phantoms. The measured carotid diameter change in the volunteers ranged from 7.5 to 11.8% (mean = 9.8%) and agrees closely with values reported in the literature. The difference between values taken one hour apart ranged from 0.2 to 0.5%. We conclude that the new system provides rapid, accurate, and repeatable measurements of vessel distension in humans.

© 2001, by The Institute of Electrical and Electronics Engineers, Inc. All rights reserved.