Transducer for Harmonic Intravascular Ultrasound Imaging

Hendrik J. Vos, Martijn E. Frijlink, Erik Droog, David E. Goertz, Gerrit Blacquière, Anton Gisolf, Nico de Jong, and Antonius F. W. van der Steen

ABSTRACT A recent study has shown the feasibility of tissue harmonic imaging (THI) using an intravascular ultrasound (IVUS) transducer. This correspondence describes the design, fabrication, and characterization of a THI-optimized piezoelectric transducer with oval aperture of 0.75 mm by 1 mm. The transducer operated at 20 MHz and 40 MHz, and was comprised of a single piezoelectric layer with additional passive layers. The Krimholtz-Leedom-Matthaei (KLM) model was used to iteratively find optimal material properties of the different layers. The transducer characterization showed -6 dB fractional bandwidths of 30% and 25%, and two-way insertion losses of -20 dB and -36 dB, respectively.

© 2005, by The Institute of Electrical and Electronics Engineers, Inc. All rights reserved.